Description

The APL-PT series automatic blowing and capturing container is mainly used for the analysis of volatile organic compounds (VOCs) in water and soil, and can also be used for the analysis of volatile compounds in food. Compliance with standards includes "HJ639-2012 Determination of Volatile Organic Compounds in Water Quality by Sweep Capture/Gas Chromatography Mass Spectrometry", "HJ686-2014 Determination of Volatile Organic Compounds in Water Quality by Sweep Capture/Gas Chromatography Mass Spectrometry", "HJ605-2011 Determination of Volatile Organic Compounds in Soil and Sediments by Sweep Capture/Gas Chromatography Mass Spectrometry", etc.

Model:PT-100
Time:2026-05-21
Nature:
Visits:5
IntroductionOnline Message

一、 Instrument features and main functions

1. Automatic leak detection and fault alarm function, easy operation, method parameter setting, real-time animation display of working status, fully automatic operation of 1-100 samples;

2. The injection valve, sample transfer tube, water trap, and capture tube desorption zone are all separately heated and temperature controlled; And output a synchronization signal during injection, which can simultaneously start the chromatography and workstation;

3. Automatic trap baking function, capable of independently running the trap baking program, baking the trap and system pipelines, and removing system residues;

4. Automatic pipeline cleaning function, capable of independently running pipeline cleaning programs for targeted cleaning;

5. The automatic trap desorption function allows for independent operation of the trap desorption injection program, facilitating the inspection of whether the trap is contaminated;

6. Internal standard addition system: Quantitative addition through 3 two-way valves on the 3-digit solenoid valve module

7. Equipped with magnetic stirring function, the sample can be automatically stirred with adjustable stirring speed;

8. The sample transfer tube and injection valve have automatic blowback function, which avoids cross contamination of different samples;

9. Liquid samples support U-tube ectopic blowing or in-situ blowing

10. All sample flow paths have undergone insulation treatment and silane inert treatment, greatly reducing analysis residues and maintaining sample integrity during operation.

11. Optional reverse control GC and GCMS, which can automatically start the dedicated startup line interface for various brands of GC and GCMS, and can connect to all models of GC and GCMS at home and abroad;

12. The sample is passed through a water trap to prevent water vapor from entering the capture tube and reduce the impact of water vapor on GC and GCMS;

二、 Main technical parameters

1. Temperature control range of the water trap: room temperature -240 ℃, with a heating power of approximately 100W set at increments of 1 ℃;

2. The temperature control range of the valve injection system is from room temperature to 220 ℃, with a heating power of approximately 200W set at increments of 1 ℃;

3. The temperature control range of the sample transfer pipeline is from room temperature to 220 ℃, with a heating power of approximately 150W set at increments of 1 ℃;

4. The capture tube adopts electronic refrigeration: -30 ℃~room temperature ℃, set in increments of 1 ℃;

5. Sample temperature control range: room temperature -240 ℃, with a heating power of approximately 100W set at increments of 1 ℃;

6. Sample bottle specification: 40ml;

7. Blow flow rate: 0-100ml/min (continuously adjustable);

8. Backflow rate: 0-100ml/min (continuously adjustable);


Online Message

Comment Box

  • Product:

  • Commpany:

  • Name:

  • Tel:

  • Email:

  • Address:

  • Message:

  • Code:

    Please enter the verification code in the filling box!

Follow the official account

Follow the official account
028-84365170

Copyright © 2026 APL Technology Group (Chengdu) Co., Ltd

Technical Support:Chemical Instrument Network    sitemap.xml