75 bit fully automatic blowing and capturing system

Description

The APL-PTC series automatic blowing and capturing container is mainly used for the analysis of volatile organic compounds (VOCs) in water and soil, and can also be used for the analysis of volatile compounds in food. Compliance with standards includes "HJ639-2012 Determination of Volatile Organic Compounds in Water Quality by Sweep Capture/Gas Chromatography Mass Spectrometry", "HJ686-2014 Determination of Volatile Organic Compounds in Water Quality by Sweep Capture/Gas Chromatography Mass Spectrometry", "HJ605-2011 Determination of Volatile Organic Compounds in Soil and Sediments by Sweep Capture/Gas Chromatography Mass Spectrometry", etc.

Model:PT-75
Time:2026-05-21
Nature:
Visits:5
IntroductionOnline Message

一、 Characteristics and main functions

1. Easy operation, method parameter setting, real-time animation display of working status, fully automatic operation of 1-75 samples;

2. The injection valve, sample transfer tube, water trap, and capture tube desorption zone are all separately heated and temperature controlled;

3. Set up the analysis program, press the run button to automatically complete sample blowing and injection, and start chromatographic analysis;

4. Through time programming, automatic functions such as blowing, trapping, desorption, injection, and backwashing cleaning can be achieved;

5. Equipped with magnetic stirring function, the sample can be automatically stirred with adjustable stirring speed;

6. The sample transfer tube and injection valve have automatic blowback function, which avoids cross contamination of different samples;

7. It can automatically start the startup line interface of various brands of GC and GCMS, and can connect to all models of GC and GCMS at home and abroad;

8. The sample passes through a water trap to prevent water vapor from entering the capture tube, reducing the impact of water vapor on GC and GCMS;

二、 Main technical parameters

1. Temperature control range of the water trap: room temperature -240 ℃, with an increment of 1 ℃. The heating power is approximately 400W;

2. Temperature control range of valve injection system: room temperature -220 ℃, with increments of 1 ℃. The heating power is approximately 60W;

3. The temperature control range of the sample transfer pipeline is from room temperature to 220 ℃, with an increment of 1 ℃. The heating power is approximately 40W;

4. The capture tube adopts electronic refrigeration, and the temperature control range for adsorption and desorption is -30 ℃ to 400 ℃, with increments of 1 ℃;

5. Sample temperature control range: room temperature -240 ℃, with an increment of 1 ℃. The heating power is set at approximately 200W;

6. Sample bottle specification: 40ml;

7. Blow flow rate: 0-100ml/min (continuously adjustable);

8. Backflow rate: 0-100ml/min (continuously adjustable);


Online Message

Comment Box

  • Product:

  • Commpany:

  • Name:

  • Tel:

  • Email:

  • Address:

  • Message:

  • Code:

    Please enter the verification code in the filling box!

Follow the official account

Follow the official account
028-84365170

Copyright © 2026 APL Technology Group (Chengdu) Co., Ltd

Technical Support:Chemical Instrument Network    sitemap.xml